Remote control mode, Remote sensing, Connection – B&K Precision 1900-1901-1902 - Manual User Manual

Page 17: 3 remote sensing

Advertising
background image

15

4.2.4 Analog Remote Control Mode

Select this mode to control the output voltage and current by Remote Control
Connector. Please refer to

4.4

Analog Remote Control

for more details.


4.3

Remote Sensing

(Model 1900 Only)

WARNING: Never short the remote sensing terminal.

Never connect the remote sensing terminal in reverse polarity.

Always disconnect the remote sensing terminal first after use.

4.3.1 Connection

1. Make the power connections between power supply and equipment.

2. Check and make sure the power connections are secure.

3. Then make connections between Remote Sense Terminal and equipment.

Figure 3

shows typical connection between power supply and device for remote

sense operation.

Figure 3 - Remote Sense Connection

The remote sensing wire gauge should be at least 22AWG.