18

Method A: Using Two External Variable DC Voltage Sources

PIN

FUNCTIONS

REMARKS

1

Internal DC +5 V

Less than 50 mA

2

Voltage Adjust

0 – 5 V

3

Current Adjust

0 – 5 V

4

Ground

5

Output OFF

Short to Ground

6

N/A

7

N/A

8

N/A

Table 4 – Remote Connector Plug Pin Assignment for External Variable Voltage Sources

A variable external DC voltage source of 0 – 5 V is fed into the analog remote control
terminal to adjust the output voltage level of both Main and Auxiliary output.

WARNING: Do not input higher than 5 V, otherwise the overvoltage protection
(OVP) will be triggered.

1. Make sure the load is disconnected and the power supply is OFF.

2. Connect pin 2 to positive polarity of first external voltage source and pin 4 to

negative polarity of first external voltage source.

3. Connect pin 3 to positive polarity of second external voltage source and pin

4 to negative polarity of second external voltage source.

4. Turn the remote control ON/OFF switch to ON position.