B&K Precision 393 - Manual User Manual

Page 23

Advertising
background image

23


AC Volts (True RMS)

Range

Resolution

Accuracy

(45 Hz - 2 kHz) / HFR2

Input

Impedance

600 mV

0.01 mV

+(1.0% rdg + 20 dgt)(1)

+(1.5% rdg + 20 dgt)(2)

10 MΩ

6 V

0.1 mV

11 MΩ

60 V

1 mV

+(1.0% rdg + 20 dgt)(1)

+(1.5% rdg + 20 dgt)(2)

+(2.0% rdg + 20 dgt)(3)

10 MΩ

600 V

10 mV

10 MΩ

750 V

100 mV

+(2.0% rdg + 20 dgt)(4)

10 MΩ

Range

Resolution

Accuracy

(45 - 60 Hz) / HFR1

Input

Impedance

600 mV

0.01 mV

+(2.0% rdg + 20 dgt)

10 MΩ

6 V

0.1 mV

11 MΩ

60 V

1 mV

10 MΩ

600 V

10 mV

10 MΩ

750 V

100 mV

10 MΩ

AC+DC Volts (True RMS)

Range

Resolution

Accuracy (45 Hz - 2 kHz)

Input

Impedance

600 mV

0.01 mV

+(1.5% rdg + 30 dgt)(1)

+(2.0% rdg + 30 dgt)(2)

10 MΩ

6 V

0.1 mV

11 MΩ

60 V

1 mV

+(1.0% rdg + 30 dgt)(1)

+(2.0% rdg + 30 dgt)(2)

+(2.5% rdg + 30 dgt)(3)

10 MΩ

600 V

10 mV

10 MΩ

750 V

100 mV

+(2.5% rdg + 30 dgt)(4)

10 MΩ

(1) For frequency range 45 Hz to 500 Hz.

(2) For frequency range 500 Hz to 1 kHz.

(3) For frequency range 1 kHz to 2 kHz.

(4) For frequency range 45 Hz to 1 kHz.

Crest factor: 3 at full scale and 6 at half scale

AC Coupled True RMS specified from 2% to 100% of range

HFR1 (High Frequency Reject): >1 kHz

HFR2 (High Frequency Reject): >100 kHz

Peak hold accuracy: +(3.0% rdg + 500 dgt) for frequency 45 Hz to

500 Hz at 60 V AC to 750V AC

Overload protection: 1000 V DC or 750 V AC RMS

Advertising
This manual is related to the following products: