Resistance, Continuity test, Diode test – B&K Precision 393 - Manual User Manual

Page 25: Conductance (6000 counts)

Advertising
background image

25

Resistance

Range

Resolution

Accuracy

Open Circuit

Voltage

600 Ω

0.01 Ω

+(0.3% rdg + 20 dgt)

-3.0 V DC typical

6 kΩ

0.1 Ω

+(0.3% rdg + 10 dgt)

-1.2 V DC typical

60 kΩ

1 Ω

+(0.3% rdg + 10 dgt)

-1.2 V DC typical

600 kΩ

10 Ω

+(0.3% rdg + 10 dgt)

-1.2 V DC typical

6 MΩ

100 Ω

+(1.0% rdg + 10 dgt)

-1.2 V DC typical

60 MΩ

1 kΩ

+(3.0% rdg + 20 dgt)

-1.2 V DC typical

Overload protection: 600 V DC or 600 V AC RMS

Continuity Test

Range

Audible

Threshold

Response

Time

Open Circuit

Voltage

600 Ω

<40 Ω

Approximately 100 ms -3.0 V DC typical

Overload protection: 600 V DC or 600 V AC RMS

Diode Test

Range/

Resolution

Test

Current

Accuracy

Open Circuit

Voltage

2 V / 1 mV

0.5 mA

typical

+(2.0% rdg + 10dgt)

3.0 V DC typical

Audible indication: Less than 0.05 V DC
Overload protection: 600 V DC or 600 V AC RMS

Conductance (6000 counts)

Range

Resolution

Accuracy

Open Circuit

Voltage

60 nS

0.01 nS

+(1.0% rdg + 10 dgt) -0.7 V DC typical

Equivalent resistance range: 16.7 MΩ to 100 GΩ

(S = Siemens = 1 / Ω )
Overload protection: 600 V DC or 600 V AC RMS

Advertising