B&K Precision 5492BGPIB - Manual User Manual

Page 76

Advertising
background image

SCPI Command Reference

76

:FORMat <name>

Command syntax:

:CALCulate2:FORMat <name>

Command Parameter:

<name> = NONE

No calculation

MAXimum Largest reading in buffer

MINimum Lowest reading in buffer

MEAN

Mean (average) value of readings in buffer

SEDViation Standard deviation of readings in buffer

Query:

:FORMat?

Query programmed math format.

Return String:

NONE, MEAN, SEDV, MAX, or MIN

Description:

This command is used to specify

which statistical calculations to recall (using “:CALC2:DATA?”

command) from the stored readings in the buffer. If NONE is selected or if the buffer is empty,
using “:CALC2:DATA?” command will return the current measured reading on the display.

:STATe <b>

Command syntax:

:CALCulate2:STATe <b>

Command Parameter:

<b> = 0 or OFF

Disable store reading

1 or ON

Enable store reading

Query:

:STATe?

Query state (ON or OFF) of store reading function

Return String:

1 or 0

Description:

This command is used to enable or disable

store reading. It is equivalent to setting “RDGS

STORE” to ON from the front panel menu. When enabled, readings on the display will immediately
begin storing into the buffer if the meter is already in measurement mode.

For USB/RS-232 Interface: To store reading subsequently with this command, the buffer must be
cleared prior to sending this command using the “:CALC2:TRAC:CLE” command.