B&K Precision 5492BGPIB - Manual User Manual
Page 90

SCPI Command Reference
90
:COUNt <n>
Command syntax:
[:SENSe[1]]:CURRent:AC:AVERage:COUNt <n>
[:SENSe[1]]:CURRent:DC:AVERage:COUNt <n>
[:SENSe[1]]:VOLTage:AC:AVERage:COUNt <n>
[:SENSe[1]]:VOLTage:DC:AVERage:COUNt <n>
[:SENSe[1]]:RESistance:AVERage:COUNt <n>
[:SENSe[1]]:FRESistance:AVERage:COUNt <n>
Command parameter:
<n> = 1 to 100
Specify filter count
MIMimum 1
MAXimum 100
Query:
:COUNt?
Query filter count
Description:
These commands are used to specify the filter count. In general, the filter count is the number of
readings that are acquired and stored in the filter buffer for the averaging calculation. The larger the
filter count, the more filtering that is performed.
THReshold Command
Use these commands to set the maximum range input (signal level) for frequency and period
measurements.
:RANGe <n>
Command syntax:
[:SENSe[1]]:FREQuency:THReshold:VOLTage:RANGe <n> Set voltage threshold range.
[:SENSe[1]]:PERiod:THReshold:VOLTage:RANGe <n>
Set voltage threshold range.
Command parameter:
<n> = 0.1 (120 mV), 1 (1.2 V), 10 (12 V), 100 (120 V), 750 (750 V)
Specify signal levels in volts (voltage threshold)
Query:
:RANGe?
Query maximum signal level.
Return String:
0.1, 1, 10, 100, or 750
Description:
This command is used to set the current or voltage threshold range for frequency and period