6 - Set up

6.1 - Self-diagnostics


Performs tests useful for maintenance staff.

6.2 - Self-calibration

For machine self-calibration proceed as follows:
- Fit a medium-sized wheel with steel rim on the shaft. Example: 6” x 14” (± 1”)
- Preset the exact dimensions of the wheel mounted.

CAUTION!!

Presetting of incorrect dimensions would mean that the machine is not correctly calibrated,

therefore all subsequent measurements will be incorrect until a new self-calibration is performed

with the correct dimensions!

- Perform a spin under normal conditions.


- Add a 100 g. sample weight (3.5 oz) on the outside in any angular position.

- Shift the sample weight from the outside to the inside keeping the same
angular position.

- Shift the sample weight to the 12 o’clock position.

END OF SELF-CALIBRATION

CANCELS SELF-CALIBRATION IN ANY PHASE.