Part 3 - keypad configuration, Basic menu units selection, Important! k-factor entry – Dynasonics MFX Series Insertion MagProbe Flow Meter User Manual
Page 37

Rev. 01/11
-3.3-
MFX
The following sections define the configuration parameters located
in each of the menus.
The BASIC menu contains all of the configuration parameters
necessary to make the MFX operational.
UNITS
ENGLSH
METRIC
Installs a global measurement standard into the operation of the
instrument. The choices are either English or Metric measurements.
x
Select ENGLSH if all configurations (pipe sizes, etc.) are to be
made in inches. Select METRIC if the meter is to be configured
in millimeters.
K FACTOR -- MagProbe K-Factor Entry
0.1 - 100.00
Each MagProbe has a unique K-factor that is factory set and
recorded on the information tag located on the MagProbe flow
sensor. To enter the MagProbe K-factor into the MFX, press the
ENTER key, then use the up/down arrow keys to adjust displayed
value to equal the unique K-factor value. When the value is proper,
press the ENTER key to record the value. This procedure must be
followed if a replacement MagProbe is procured from the
Dynasonics factory.
PART 3 - KEYPAD CONFIGURATION
1. BASIC MENU
UNITS
Selection
NOTE: If the UNITS entry has been changed from ENGLSH to
METRIC or from METRIC to ENGLSH, the entry must be saved
and the instrument reset (power cycled or System Reset entered)
in order for the MFX to initiate the change in operating units.
Failure to save and reset the instrument may result in meter not
measuring properly.
IMPORTANT!
K-factor Entry