Section 1 Introduction

Brooks

®

Digital MFC's & MFM's

Installation and Operation Manual
X-TMF-SLA5800-MFC-eng
Part Number: 541B027AAG
April, 2013

DeviceNet Communications
The Brooks SLA5800 Digital Series is also available with DeviceNet

TM

communication capability. DeviceNet is an open digital protocol capable of
high speeds and easy system connectivity. Brooks Instrument has several
of its devices available on this popular networking standard, and is a
member of ODVA

TM

(Open DeviceNet Vendors Association), the governing

standard body for DeviceNet.
DeviceNet

is similar to the RS485 standard in that it is a multi-drop

connection that allows a maximum of 64 devices to be connected on the
same network. Baud rate selections for DeviceNet

products are 125K,

250K and 500K and can be selected via MAC ID switches mounted on the
device.
The DeviceNet

communication link also provides access to many of the

Brooks SLAMf Digital Series functions for “control and monitor” operations,
including:

Accurate setpoint adjustment and flow output measurement (including
units of measure selection)

PID Settings (controller only)

Valve Override (controller only)

Calibration Gas Select

Soft Start Control (controller only)

F

OUNDATION

®

Fieldbus Communications:

The Brooks SLA5800 Digital Series is supporting F

OUNDATION

®

Fieldbus

communication protocol. F

OUNDATION

®

Fieldbus is a digital network allowing

usage of existing 4-20mA cables, avoiding costly re-wiring. Fully certified
by passing ITK, this device has passed several Interoperability
requirements over a broad range of hosts. When combined with DeltaV
and using the power of PlantWeb, those devices provide intelligent alerts
allowing accurate device maintenance and service.

Value Range check - Part of the standard function blocks

Temperature sensor connection - Check sensor connection

Firmware checksum - Check for Internal firmware integrity

Non-volatile memory - Check for non-volatile memory integrity

RAM - Check for RAM integrity

Zero Drift/Valve Leak-by - Check for flow leak-by or sensor zero drift

Device Overhaul due - Preventive Maintenance

Calibration Due - Preventive Maintenance

Valve spring life - Preventive Maintenance

No Flow - No flow detected when setpoint requested

Reverse Flow - Reverse flow detected

·

Flow Totalizer - Informed when a user define amount of fluid has

been delivered

·

Time Totalizer - Informed when a user define amount of time has

expired

Device type dependant function block are available representing the
different device functions:

• Current Flow Value (Mass Flow device only)
• Current Pressure Value (Pressure device only)
• Current Device Temperature (Mass Flow device only)
• Current Valve position (Controller Only)
• Setpoint Control (Controller Only)
• Direct Valve Control (Controller Only)
• Actuator Override (Controller Only)
• Ultra-fast (8ms) PID function block for Cascade control

(all devices)