Burkert Type 1150 User Manual
Page 79
Advertising

1150 - 77
C
ONFIGURATION
Value / selection
Description
Sensor voltage
(Analog selector)
Switched off
Source for the voltage signal of the zircon dioxide sensor.
Sensor temperature
(Analog selector)
Switched off
Source for the temperature signal of the zircon dioxide
CO measurement
(Analog selector)
Switched off
Source for the measurement signal of the CO content
CO content
0 to 30 to 9999
If the CO content is not measured, a fixed value can be defined here
Correction value
0 to 1 to 9999
The correction value is a C-level determined through reference measurement
Cycle time
0 to 9999 min.
Cycle time for sensor flushing
Flushing time
0 to 9999 min.
Flushing time for sensor flushing
Recovery time
0 to 9999 min.
Recovery time for sensor flushing