Calibrating the flow zero point – Burkert Type 8045 User Manual

Page 59

Advertising
background image

55

Operatingandfunctions

8.7.2.

calibrating the flow zero point

Adjust this parameter:

• before carrying out a teach-in procedure of the K factor.

• after maintenance work.

• if the measured flow rate is not zero whereas the fluid circulation has been stopped.

Make sure there are no bubbles in the pipe.

Before calibrating the flow zero point:

• at first commissioning the device, immerse the sensor in fluid for 24 hours before calibration.

• after maintenance work, immerse the sensor in the fluid for 1 hour before calibration.

During the calibration:

• the device status LED flashes.

• the outputs are frozen to the last measured flow rate value.

• the device cannot be configured.

The flow zero point calibration can be:

• either remotely triggered via the digital input. See chap. “8.6.12” and Fig. 56.

• either carried out using this parameter in the Test menu; See Fig. 57.

0.000 L/s

→

Charge the pipe.

→

Stop the fluid circulation.

→

Make sure the fluid is still.

→

Make sure the device status LED is green.

→

Make sure the device is in the Process level and not in the Configuration level.

→

Activate the digital input configured to trigger the flow zeo point calibration.

mEAsURE /

The device calibrates the flow zero point during approximately
30 seconds.

• if the device status LED remains green, the calibration of the flow zero point has

succeeded.

• if the device status LED is orange, the calibration of the flow zero point failed: the

message "CAL. FAIL" is displayed in the warning message list. Restart the calibration.

Process level

Fig. 56:

Flow zero point calibration via digital input

English

Type 8045