6 calibration, 1 temperature sensor calibration, 5 sensor setup menu (continued) – Hach-Lange 9187 sc User Manual

Page 23: Operation

Advertising
background image

21

Operation

4.6 Calibration

4.6.1 Temperature Sensor Calibration

The probe contains a temperature sensor which is factory pre-set. If there are setting
questions, the data can be validated using a high precision thermometer using

Table 1

and performing the steps in

section 4.6.1.1 on page 22

.

DISP PH FORMAT (9184sc T.F.C or 9184sc Chlorine + Acid only)

Choose either XX.XX pH or XX.X pH.

pH MAXIMUM (9184sc T.F.C only)

Allows user to set the maximum pH allowed value. An higher value will display a PH TOO HIGH error message.

LOG SETUP

Allows user to select data logging interval for the sensor and temperature.

FILTER

Adjust for + s. This averages the signal over the specified time interval.

MAINS FREQ

Choose 50 or 60 Hz.

CONFIGURE (continued)

DEFAULT SETUP

Resets the sensor configuration to default settings.

DIAG/TEST

PROBE INFO

Displays the driver and software versions and the serial number.

CAL DATA

Displays OFFSET: °C, SLOPE: in A/mg and OFFSET: uA, SLOPE: %

SIGNALS

Displays INT, TEMP RAW, MV RAW and PH RAW.

COUNTERS

Displays sensor total time and humidity suppressor.

4.5 Sensor Setup Menu (continued)