Section 6

Troubleshooting

6.1 Error Messages

Note: When an error occurs, the measurement values are replaced by dashes, (- - -).

6.2 Warning Messages

Message Type

Error Message

Solution

Measurement-related

Error Messages

CONC TOO HIGH

Check the current value, along with the calibration parameters. Check
electrode.

CONC TOO LOW

Check the current value, along with the calibration parameters. Check
electrode.

TEMPERATURE ERROR

Check for short-circuit or open circuit.

INT. TOO LOW

Negative current. Check the electrode (electrolyte and membrane).

INT. TOO HIGH

Ensure that there are no short-circuits on the measurement chain.
Check the polarizing voltage.

***** on the main display

No communication. Verify the connection and the cable. Test the 12V
power supply.

(At Connection time:) SENSOR

MISSING

FFFFFFFFFFFFF displays

No communication. Verify the sensor is properly connected to the
transmitter. Verify that the cable is not damaged. Test the 12V power
supply. Open the sensor and change the board.

COMMUNICATION ERROR

Open the sensor and verify that there is no humidity.

TEMP TOO LOW

Temperature measurement is below –2°C. Verify that the actual
temperature is not below –2°C. Verify the internal resistance of the
NTC/K, which must be around 10 K. Connect the sensor simulator and
verify the RAW value.

TEMP TOO HIGH

Verify the actual temperature is not higher than 45 °C. Connect the
sensor simulator and verify the RAW value.

RAW MEASURMENT

Change the preamplifier.

PH TOO LOW

(9184sc only)

PH electrode is clogged, broken or too old. Calibrate the pH electrode.
Clean the electrode. Replace the electrode.

PH TOO HIGH

(9184sc only)

PH electrode is clogged, broken or too old. Calibrate the pH electrode.
Clean the electrode. Replace the electrode.

Calibration-related

Error Messages

ΔT OUT OF LIMITS

The temperature difference between calibration and the theoretical
sensor response is greater than the allowed limit. Limits: ±20 °C. Check
temperature calibration (see

4.6.1 Temperature Sensor Calibration

),

OUT OF 4/20 mA

The measured value is out of the programmed scale range for analog
outputs 1 and 2.

Warning Display

Problem

Resolution

CAL FAIL SLOPE

LOW

Slope is outside the limits.

Adjust so that it is within the limits by checking zero cal, ensuring the
proper flow rate, and validating that it’s clean. Be sure to put in the
actual value and not the offset value when adjusting.

CAL FAIL SLOPE

HIGH

Adjust so that it is within the limits by checking zero cal, ensuring the
proper flow rate, and validating that it’s clean. Be sure to put in the
actual value and not the offset value when adjusting.

CAL TOO OLD

The last calibration was more
than x days ago. (Setting from
the sensor setup)

Calibrate the sensor.

Set the calibration interval in the sensor setup.