Section 1 specifications – Hach-Lange 9187 sc User Manual

Page 5

Advertising
background image

3

Section 1

Specifications

Specifications are subject to change without notice

General

Mounting

Flat, vertical surface such as a wall, panel, stand, etc.

Analyzer Dimensions

10.63 x 9.84 in. (270 x 250 mm)

Analyzer Weight

6.5 kg (14.3 lb)

Materials

Electrode: gold cathode/silver anode; probe body: PVC; measuring cell: acrylic

Sample Requirements

Sample Flow Rate to Analyzer Flow rate should allow for 14 L/hr minimum
Minimum Inlet Pressure to
Instrument

0.1–2 bar (1.4–28 psi)

Minimum Flow Rate

14 L/hr auto-regulated by flow thru cell

Pressure Range

0.1–2 bar (1.4–28 psi) influent; flow cell pressure will be the atmospheric pressure

Sample Temperature Range

+2–45 °C (35.6–113 °F)

Temperature Compensation

Automatic over the sample Temperature Range

Sample pH

4 to 8 (acidification unit available for >8 pH)

Sample Inlet Tubing: at
instrument

¼ in. OD

Drain Fitting

½

in. ID (supplied)

Application Sample

Clean water

Electrical

Power Consumption

12 V, 1.5 Watts provided by sc controller

Performance

9184sc

9185sc

9187sc

Measurement Range

0–20 ppm (0–20 mg/L) HOCl

0–2 ppm (0–2 mg/L) O

3

0–2 ppm (0–2 mg/L) ClO

2

Detection Limit

5 ppb (0.005 mg/L) HOCl

5 ppb (0.005 mg/L) O

3

10 ppb (0.01 mg/L) ClO

2

Accuracy

2 % or ±10 ppb HOCl,
whichever is greater

3 % or ±10 ppb O

3

,

whichever is greater

5 % or ±10 ppb ClO

2

,

whichever is greater

Standard Deviation

0.7 %

1.0 %

1.5 %

Interference

No interference from
chloramines.
Ozone and chlorine dioxide
interfere with measurement.

No interference from
chlorine, chloramines,
hydrogen peroxide, bromine,
or chlorine dioxide.

None

Repeatability

±10 ppb (0.01 mg/L) or ±5 %, whichever is greater at a pH < 7.5

Response Time

90 % < T=90 seconds

Measurement Interval

Continuous

Measurement Technology

Amperometric/Membrane (electrode, membrane, electrolyte)

Calibration

Electrical zero or chemical zero with dechlorinated or deozoned water; calibration of the
slope by comparison with a laboratory instrument; pH calibration (9184sc only) with Single or
Two Point using standards or comparison with lab method with the process sample.

Calibration Interval

2 months for typical application

Maintenance

Maintenance Interval,
Measurement Cell

6 months for the membrane and electrolyte for typical operation (3 to 12 month range)

Maintenance Interval, pH

1 to 1.5 years for typical operation