9184sc Theory of Operation

It is also important to notice that the dissociation constants are temperature-dependent
(the equipment takes into account this element).

The amperometric sensor consists of:

a gold working electrode (cathode) where the main reaction occurs

a silver counter-reference electrode (anode)

KCl electrolyte

a microporous membrane selective to HOCl

The HOCl molecules in the sample diffuse through the membrane to a thin region between
the membrane and the cathode that contains the electrolyte.

A constant potential is applied to the working electrode where HOCl is reduced according
to the reaction:

At the silver electrode (anode) the silver is oxidized to Ag

+

:

The reduction in HOCl at the cathode generates a current that is directly proportional to its
partial pressure in the sample.

The electrochemical reaction and diffusion through the membrane are dependent upon
temperature. Consequently, the measurement cell contains a temperature sensor that
allows for automatic temperature compensation.

In the acidification version an additional analyzer version offers the possibility to measure
samples with a high pH value. The sample pH is kept constant between 5.5 and 6.5 by
continuously adding acid solution. At such pH levels, all ClO

ions turn into HOCl, thus

enabling the sensor to measure TFC.

HOCl H

+

2e

Cl

H

2

O

+

+

+

2Cl

2Ag

+

+

2AgCl 2e

+