W A R N I N G

Electrical shock and fire hazards. Make sure that the supplied cord and non-
locking plug meet the applicable country code requirements.

W A R N I N G

Electrocution hazard. Make sure that the protective earth conductor
has a low impedance connection of less than 0.1 ohm. The connected
wire conductor must have the same current rating as the AC mains line
conductor.

N O T I C E

The instrument is used for a single phase connection only.

Cord installation: The manufacturer recommends to use the optional
cord and sealing gland. Refer to the maintenance manual for the
replacement parts list. For a customer-supplied cord, three 18-gauge
conductors are required and the cord must be shorter than 3 meters
(10 feet). Use a sealing type strain relief to keep the environmental rating
of the instrument. Refer to

Specifications

on page 3. To connect power

to the instrument, refer to

Table 3

or

Table 4

and

Figure 13

.

Table 3 AC wiring information (AC models only)

Terminal

Description

Color—North
America

Color—EU

1

Protective Earth (PE)
Ground

Green

Green with
yellow stripe

2

Neutral (N)

White

Blue

3

Hot (L1)

Black

Brown

Table 4 DC wiring information (DC models only)

Terminal

Description

Color—North
America

Color—EU

1

Protective Earth (PE)
Ground

Green

Green with
yellow stripe

2

24 VDC return (–)

Black

Black

3

24 VDC (+)

Red

Red

14 English