Connect to the relays, Connect to the 4–20 ma outputs, Figure 14 – Hach-Lange POLYMETRON 9610 sc SiO2 Operations User Manual

Page 16

Advertising
background image

Figure 14 Device connection

Connect to the relays

D A N G E R

Electrocution hazard. Do not mix high and low voltage. Make sure that
the relay connections are all high voltage AC or all low voltage DC.

C A U T I O N

Fire hazard. Relay loads must be resistive. Always limit current to the
relays with an external fuse or breaker. Obey the relay ratings in the
Specifications section.

C A U T I O N

Chemical exposure hazard. Obey laboratory safety procedures and
wear all of the personal protective equipment appropriate to the
chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.

N O T I C E

Wire gauge less than 18 AWG is not recommended.

The analyzer contains relays for sample concentration alarms (2x),
analyzer system warning and analyzer system shutdown. Refer to

Wiring

connections overview

on page 13 to connect a device (NO = normally

open, COM = common, NC = normally closed).

Connect to the 4–20 mA outputs
Use twisted pair shielded wire for the 4–20 mA output connections.
Connect the shield at the recorder end or the analyzer end. Do not
connect the shield at both ends of the cable. Use of non-shielded cable
can result in radio frequency emission or susceptibility levels higher than
the allowed levels.
Refer to

Wiring connections overview

on page 13 to connect the device.

Refer to

Specifications

on page 3 for wiring and load impedance

specifications.

Note: The 4-20 mA outputs cannot be used to provide power to a 2-wire (loop-
powered) transmitter.