Plumb the sample and drain lines, Figure 5 – Hach-Lange POLYMETRON 9610 sc SiO2 Operations User Manual

Page 9

Advertising
background image

Figure 5 Ports for six sample streams

1 Sample 6 bypass

drain

6 Sample 2 bypass

drain

11 Sample 2 inlet

2 Sample 5 bypass

drain

7 Sample 1 bypass

drain

12 Air purge inlet

(optional)

3 Sample 4 bypass

drain

8 Chemical drain

13 Sample 3 inlet

4 Sample 6 inlet

9 Case drain for spills

or leaks

14 Sample 4 inlet

5 Sample 3 bypass

drain

10 Sample 1 inlet

15 Sample 5 inlet

Plumb the sample and drain lines

C A U T I O N

Explosion hazard. Use only the supplied regulator from the
manufacturer.

C A U T I O N

Chemical exposure hazard. Dispose of chemicals and wastes in
accordance with local, regional and national regulations.

N O T I C E

Do not connect the drain lines to other lines or backpressure and damage to the
analyzer can occur. Make sure that the drain lines are open to air.

N O T I C E

To prevent backpressure and damage to the analyzer, make sure that the
analyzer is higher than the facility drain(s) used and that the drain line has a
constant downward slope.

N O T I C E

The pressure regulator is set to a fixed pressure and cannot be changed.

Use the supplied tubing (6 mm), Y-strainer with filter and pressure
regulator to plumb the drain and the sample to the analyzer. Refer to

Figure 6

. The sample line tubing that goes into the plumbing access

ports must be 6 mm. Tubing of 1/4 in. may be used for the sample line
up to the valve/y-strainer but not into the plumbing access ports of the
analyzer.

English 9