Troubleshooting – Hach-Lange LICO 690 User Manual User Manual

Page 104

Advertising
background image

104

Troubleshooting

Fault
Test program stopped!
Please check lamp
Close the lid.
Error [xx]

Test program stops when the instrument
is started

Check the lamp and replace it if
necessary.
Close lid.
Press Start Again.

Fault
Test program stopped!
Please remove the cuvette
Close the lid.

Test program stops when the instrument
is started

Remove the cuvette/sample cell from
the cell compartment.
Press OK.

Error
Selfcheck stopped.
Hardware error.
Error [x]

Electronic defect

Contact the manufacturer or national
agent and specify the error number.

Error
Too much ambient light!
Move instrument into shade
or close the lid

The instrument sensors detects too
much ambient light.

Decrease ambient light. Avoid direct
sunlight.
Close lid.

No instrument backup present!

Check the USB memory stick.

No valid data for these parameters!

Data analysis not possible, no
measurement data

Change the selection.

No valid data found!

Data cannot be viewed in the data log

Change the selection.

No measurement data present!

Data analysis settings cannot be
configured without measurement data.

Change the selection.

Control range not reached!

Data analysis limits not reached

This is a warning notice. The control
limit set was not reached.

Control range exceeded!

Data analysis limits exceeded.

This is a warning notice. The control
limit was exceeded.

Concentration too high!

Calculated concentration is higher than
999999

Dilute sample and measure again

Possible interference by:

Interference Check

The analysis is possibly erroneous due
to interferences.

Possible interference from:

Interference Check

The analysis is possibly erroneous due
to interferences.

Next service is due!

Contact the manufacturer or national
agent for an instrument service. For
setting the service reminder, refer to

section 6.1.7, page 93

Negative result!

The calculated result is negative

Check concentration of sample

Network switched off.

Network setup off, when access to
homepage via sidebar

Activate the online connection.

Remote server cannot be reached.

Error during network setup

Make sure that the instrument is
connected to the network.

Unstable lighting conditions!

Avoid direct sunlight at the measuring
location.

Please insert adapter Z.

For measurements with 11 mm round
cuvettes, adapter Z is required.

Insert adapter Z into cell compartment
#2. Confirm with OK.

Insufficient memory for update .

Error during update.

Select a memory with more space.

Error displayed

Definition

Resolution