Advanced Operations

3. Press Edit.

An automatic menu guidance queries values (filters,
wavelength, nominal values and tolerances) given in the
quality control certificate, to the following specifications:

Stray light

Photometrical accuracy

Wavelength accuracy

4. Press OK when all values are entered and the overview is

displayed.

5. Press Verification.

6. Insert cuvette adapter Z (

Figure 3, page 18

) into the cuvette

shaft (2).

7. Remove any present cuvettes from the cuvette

compartment.

8. Close the cuvette compartment cover.

9. Press Start.

10. Follow the instructions on the screen and insert the different

filters one after another in the sequence shown.

11. Press Next after inserting a filter.