Advanced Operations
3. Press Edit.
An automatic menu guidance queries values (filters,
wavelength, nominal values and tolerances) given in the
quality control certificate, to the following specifications:
•
Stray light
•
Photometrical accuracy
•
Wavelength accuracy
4. Press OK when all values are entered and the overview is
displayed.
5. Press Verification.
6. Insert cuvette adapter Z (
shaft (2).
7. Remove any present cuvettes from the cuvette
compartment.
8. Close the cuvette compartment cover.
9. Press Start.
10. Follow the instructions on the screen and insert the different
filters one after another in the sequence shown.
11. Press Next after inserting a filter.