2 software update with network connection, 3 optical checks, Advanced operations – Hach-Lange LICO 690 User Manual User Manual

Page 88

Advertising
background image

88

Advanced Operations

1. Go to http://www.hach-lange.com.

2. Select the country.

3. Enter LICO into the "Search" field.

4. Select Documents and Software from the field "according

to".

5. Navigate to the corresponding file for download.

6. Save the file to a USB storage device or PC.

7. Tap on Instrument Update in the "System Checks" menu.

8. Plug the USB storage device into the USB port (USB A) of

the device, refer to

section 3.5, page 16

.

9. Press OK. The link is established automatically and the

software is updated.

10. Press OK to return to the "System Check" menu.

Note: After the instrument update, a prompt to restart the instrument will
be displayed.

6.1.2.2

Software update with network connection

If the instrument, as described under

section 5.2.6.4

, is

connected to the Internet, the instrument will check during the
test program whether a new software update is available on the
website http://www.hach-lange.com.

1. Press OK to confirm the one-time window "Run software

update".

The link is established automatically and the software is
updated.

2. Press OK to return to the "System Check" menu.

Note: After the instrument update, a prompt to restart the instrument will
be displayed.

6.1.3

Optical Checks

The Verification Kit (refer to

Section 9, page 107

) is designed for

periodic monitoring of scattered light, photometric accuracy and
the wavelength accuracy of the spectrophotometers.

When results exceed allowable tolerances (given in the quality
control certificate to the test record), contact the manufacturer.

1. Press Optical Checks in the "System Check" menu.

2. Press Nominal Values.

Advertising