To use the 10 mm square cuvette and 11 mm round cuvettes, adapter Z
must be inserted into the cuvette compartment (2). For measurements
with 50 mm square cuvettes, you must remove the adapter.

Take a color measurement

1.

Press Color Measurement.

2.

Insert a cuvette/sample cell with distilled water to calibrate.

Note: Always carry out the calibration very carefully, as a faulty
calibration can cause inaccurate results to be obtained.

3.

The calibration starts automatically once the instrument has

detected the cuvette.
The type of cuvette/sample cell used and the exact progress of
calibration is shown in a separate window.

4.

After calibration, the cuvette size used is displayed in the top right.
Note: After calibration, you can measure the cuvette with distilled
water again as a sample.
The displayed measured values should match the uncolored color
indices (i.e. Hazen = 0, Gardner = 0.0, CIE L*a*b* = 100.0, 0.0, 0.0,
etc.)