Error when calling up the local IP address.

Network setup: DHCP client has no connection to the
DHCP server

Enter the IP address again.

Error during default gateway setup.

Network setup: default gateway cannot be set for fixed
IP address

Try to create the connection again.

Error during network drive setup!

Error during network setup

Check the settings.

Error during subnet mask setup.

Network setup: Subnet mask cannot be set for fixed IP
address

Enter the subnet mask again.

Error copying from USB memory stick.

Error during update

Restart the process or contact the manufacturer or
national agent.

Error in FTP connection.

FTP error

Make sure that the instrument is connected to the
network.

Fault
Test program stopped!
Please check lamp
Close the lid.
Error [xx]

Test program stops when the instrument is started

Check the lamp and replace it if necessary.
Close lid.
Press Start Again.

Fault
Test program stopped!
Please remove the cuvette
Close the lid.

Test program stops when the instrument is started

Remove the cuvette/sample cell from the cell
compartment.
Press OK.

Error
Selfcheck stopped.
Hardware error.
Error [x]

Electronic defect

Contact the manufacturer or national agent and specify
the error number.

Error
Too much ambient light!
Move instrument into shade
or close the lid

The instrument sensors detects too much ambient light.

Decrease ambient light. Avoid direct sunlight.
Close lid.

No instrument backup present!

Check the USB memory stick.

No valid data for these parameters!

Data analysis not possible, no measurement data

Change the selection.

Error displayed

Definition