Troubleshooting – Hach-Lange LICO 690 Basic User Manual User Manual

Page 58

Advertising
background image

English 58

Troubleshooting

Error displayed

Definition

Resolution

An error occurred when uploading the instrument
data.

Restart the process or contact the manufacturer or
national agent.

An error occurred when reading from the USB
memory stick.

Restart the process or contact the manufacturer or
national agent.

An error occurred when writing to the USB memory
stick.

Restart the process or contact the manufacturer or
national agent.

Please check on the current update file.

Error during update.

Check the USB memory stick.

Please contact Customer Service.

Error during update.

Contact the manufacturer or national agent.

Please check network configuration.

Check the network setup. .

Please check the connection.

Check the network setup. .

Please close the cover.

Close the cuvette compartment cover.

Please insert the USB memory stick.

Insert a USB memory stick into a USB A port on the
instrument.

Please check the connection and contact the
administrator.

Network setup or FTP error

Check the network setup or contact the manufacturer
or national agent.

File for instrument update missing.

Error during update.

Check the USB memory stick.

File for instrument update is faulty.

Error during update.

Save the update file again and repeat the procedure.

It’s recommended to execute a Full System Check

Check of the air values failed

Switch the instrument off and then back on again. If the
system test is unsuccessful, contact the manufacturer
or national agent.

Entry invalid!

Password incorrect

Forgotten your password?
Contact the manufacturer or national agent

Absorbance > 3.5!

The measured absorbance exceeds 3.5

Dilute sample and measure again

Color = ***

Color value outside measurement range.

Dilute the sample or select appropriate color scale.

Advertising
This manual is related to the following products: