3. Installation

3.3. Cell cables

Compensation of the cable length
To minimize the effect of the cable resistance in a conductivity measurement, the cable length is compensa-
ted by the 4-wire system. With the Pt 1000 temperature sensor, the cable resistance has little importance.
However, in order to reduce the resistance of the conductor to a half, 2 cables have been connected to
the ends of the sensor.

3.3.1. Cells with MP-5 connector

Notas importantes:
• El si628 C does not require ground connec-

tion (terminal 1).

• Connect the ground only when there could

be interference problems.

DANGER

Relays:

- Limit the current up to 3 A by using

one fuse.

-

Do not connect cables with voltage
higher than 24 V.

Terminal assignment (rear panel)

Transmitter si628 C

RED

VCA

CELL

4-20 mA

L 1

L 2

S/N:

CRISON INSTRUMENTS, S.A.
E-08328 ALELLA - Barcelona

Made in Spain

IP 54

TERMINALS

COLOUR

si628 C

(wall mount)

si628 C

(panel mount)

6

16

White

6

16

Brown

7

15

Pink

7

15

Yellow

8

14

Shield

9

13

Green

10

12

Blue

11

11

Gray

12

10

Red

MP-5 connector, 5 contacts