4. Operation

4.1. Description

Display

si628 C have an alphanumeric display that explains itself. One of the most significant screens is des-
cribed below as an example:

The instrument is already measuring conductivity,
with the following configuration:
• Temperature coefficient, CT, fixed at 2% / °C.
• Cell constant C=1 cm

-1

.

• Reference temperature, RT, fixed at 25°C.
• Limit relay is activated at 1000 µS/cm, and

desactivated at 800 µS/cm.

• Alarm relay is activated after 1minute with the

reading out of limit.

• Output current 4 mA = 200 µS/cm, 20 mA =

2000 µS/cm.

To modify the default configuration (except from
the temperature Coefficient and the reference tem-
perature) see page 16.

4.2. Setting up

Connect the si628 C to mains. At power up the first time only the following screens will appear:

Keys

ENTER.
• Horizontal movements to move forward in the “diagram”.
• Acceptance of numerical values.

ESCAPE.
• Horizontal movements to move back in the “diagram”.
• Vertical movements to move back in the “diagram”.
• To erase numerical values.

• Modification of numerical values.
• Cursor “>” movements on the screen.

si628_C

V1.0

Français

English

Deutsch

Português

LANGUAGE >Español

Italiano

MEASURE S/cm

MEASURING 22.3°C

545 µS/cm @25°

Possibility to select other
language with

keys.

MEASURING 23.2

°C

538 µS/cm

@25 °C

Sample temperature
Resolution:
1 °C manual selection.
0.1 °C measuring with Pt 1000

Measured value.

Reference temperature

Measuring units.