Programming, Operation – Hach-Lange SI628 C User Manual

Page 20

Advertising
background image

16

4. Operation

4.4. Programming

The instrument is supplied with the default configuration specified in Setting up, page 14.
To modify the default configuration and adapt it to specific applications follow the diagram below.

Select the nominal value of cell
constant to be used. If the exact
value is known, it can be directly
entered, avoiding the calibration
procedure.

Selection of the functioning
values of the corresponding
relays.
Input “ON” (closed contact).
Output “OFF” (open contact).
Selectable values between
0.0 µS and 200 mS.

Select the time before the alarm
activation.
Selectable between 1 to 99 min
or hours. 0= deactivated.

Select the conductivity values
corresponding to the Current
Output
(4 mA and 20 mA).
Selectable values between
0.0 µS and 200 mS.

Selection of the current output
when the instrument is in Hold
state: fixed at 21 mA or at the
mA corresponding to the last
measured value.

Selection of the working
language: English, German,
Spanish, French, Italian,
Portuguese.

(*) LIMIT and OUTPUT 4-20 mA:
Allows the introduction of five
digits, placing the decimal point
you want.

Note: When programming, the instrument is
in hold state:
• The contact relays are open.

The current output is frozen at 21 mA.

To quit the hold state it is necessary to return
to MEASURE.

Vary the underlined digit with

and accept with

Vary the underlined digit with

and accept with

CELL

CALIBRATE

CODE 000

CONST:
1.000 cm

–1

LIMIT (*)

ON - - - - - - OFF

1000 µS 800 µS

LANGUAGE

LANGUAGE > Español

Italiano

MEASURE S/cm

MEASURING 22.3°C

545 µS/cm @25°

OUTPUT 4-20 mA

4 mA 20 mA

pH: 0.0 2.0 mS

MEASURING 22.3°C

545 µS/cm @25°

Enter code 100.

STORING DATA

HOLD

ALARM

> 21 mA

last

ALARM:
01 min

Vary with

Vary with

si628 C_HACH LANGE_Ed. 0510.indd ANG:16

si628 C_HACH LANGE_Ed. 0510.indd ANG:16

20/5/10 11:08:45

20/5/10 11:08:45