3. Installation

ENGLISH

3.4. Protective wiring of switching contacts

Relay contacts are subjected to electrical erosion.
Especially with inductive and capacitive loads, the service life of the contacts will be reduced.
For suppression of sparks and arcing, components such as RC combinations, nonlinear resistors, series
resistors and diodes should be used.

Important note:
Fit a jumper between terminals 9 and 10 (12 and 13 in panel mount instrument) and another between
the terminals 11 and 12 (10 and 11 in panel mount instrument).

3.3.2. Cells with connector S8

Typical protective wirings

AC applications with inductive load

DC applications with
inductive load

AC/DC with
capacitive load

Load

Contact
relay

Diode
1N4007

Typical RC at 230 VAC: Capacitor 0.1 µF/630 V

Resistor 100 Ω/1 W

RC

RC

R

Load

R, for example:
5 Ω / 1 W at
24 V / 0.4 A

Maximum relays load: AC < 24 V / < 3 A / < 72 VA.

TERMINALS

COLOUR

si628 C

(wall mount)

si628 C

(panel mount)

9

13

Middle

10

12

Middle

11

11

Shield

12

10

Shield

si628 C_HACH LANGE_Ed. 0510.indd ANG:13

si628 C_HACH LANGE_Ed. 0510.indd ANG:13

20/5/10 11:08:43

20/5/10 11:08:43