11

3. Installation

ENGLISH

When making any wiring connections to the instrument, any warnings and notes found throughout the
manual must be adhered to. For more safety information see “Safety information” on page 5.

3.2.2. Connections

System functions

DANGER

Always disconnect power to the instrument when making any electrical connections.

1 Ground

connection,

see

Important

notes (page 12).

2,3

Power supply 230 or 24 VCA ± 10%.

(115 VCA upon request).

4,5

Alarm relay contacts.

6,7

Limit 1 relay contacts.

8,9

Limit 2 or Cleaning relay contacts.

10,11

Current output 4-20 mA (EC).

12 to 15 Input for 2 pole cell.
16

Cell cable shielding.

17,18

Input for temperature probe.

DANGER

High-voltage wiring for the instrument is protected by a protective casing. The protective casing
must remain in place unless a qualified installation technician is installing wiring for power.