4.4. Configuration

Note: During Configuration the instrument is in
Hold state:
• Limit and cleaning relays contact are open.

(Alarm OFF).

• Current output is frozen at the value program-

med in Hold.

To quit the Hold state it is necessary to return
to Measure. The Current output 4-20 and the
Relays will activate in 10

4.5. Cell constant selection

Select the constant cell before
modifying the configuration.

STORING

DATA

CELL

Select the nominal value of the
cell in use
. If the exact value
is known, introduce it and no
need to calibrate.

Vary with

and accept

with

CELL
1.000 cm

-1

MEASURING 22.3°C

545µS/cm @25°C

CODE 000

Enter code 100 and y accept with

Press

until the desired option is reached.

MEASURING 22.3°C

545µS/cm @25°C

MEASURING 22.3°C

545µS/cm @25°C

Conf.

> Ref. Temp.

T.C.
Relay

2

Type

measurement

CONFIGURE

CODE 000

Ref. Temp. > 25 °C
20

°C

Relay 2 > Cleaning
Limit

2

MEASURE

STORING DATA

T.C. 2.00%

LIMITS

Press

until the display MEASURE is reached.

Enter code 100 and y accept with

Press

until the desired option is reached.

Follow the diagram for
modifying the configuration.

> Conductivity (*)

Salinity