4.9. Conductivity readjustment

The si629 C allows readjusting the conductivity measured
value by modifying the reading that appears on the dis-
play. Previously the conductivity of the sample must have
been measured with a perfectly calibrated instrument.
During the readjustment the si629 C behaves as it was
calibrated with one buffer only.
Every time the si629 C is calibrated in the conventional
manner, the readjustment is lost.

or

or

+

+

Important: It is not possible to perform salinity readjustment.

4. Operation

INDIRECT CAL.

12.15 mS

INDIRECT CAL.

12.15 mS

MEASURING 18.3°C
12.15 mS/cm
@25°

Press for more than 3 sec.

P: 12.28mS

C: 0.700 cm

-1

MEASURING 18.3°C

12.28 mS/cm @25