Access codes, Operation – Hach-Lange SI629 C User Manual

Page 30

Advertising
background image

26

4. Operation

4.12. Access codes

The code 100 is for access to the calibration and programming of the instrument.
There are codes with other functions:

Code

Action

100

Main Menu.

401

Modification of the relay lag of the limits relays activation and deactivation.
Standard value: 2s. Programmable values: 0 to 99 s

413

Modification of the time for alarm activation when there is out of range reading.
Standard value: 60 s. Programmable values: 0 to 99 s/min/h

473

Modification of the delay time activation of the analogical output 4 to 20 mA after quitting Hold.
Standard value: 10 s. Programmable values: 0 to 99 s

497

Possibility to adjust the temperature for minimizing the effect of long cables.