High-voltage barrier, Wiring for power, High-voltage barrier wiring for power – Hach-Lange POLYMETRON 9523 User Manual User Manual

Page 13

Advertising
background image

Figure 4 Wiring connections overview

1 Service cable connection

5 AC and DC power connector

1

9 Discrete input wiring connector

1

2 4-20 mA output

1

6 Ground terminals

10 Digital sensor connector

1

3 Sensor module connector

7 Relay connections

1

4 Communication module

connector (e.g., Modbus,
Profibus, HART, optional
4-20 mA module, etc.)

8 Digital sensor connector

1

1

Terminals can be removed for improved access.

High-voltage barrier

High-voltage wiring for the controller is located behind the high-voltage barrier in the controller
enclosure. The barrier must remain in place except when installing modules or when a qualified
installation technician is wiring for power, alarms, outputs or relays. Do not remove the barrier while
power is applied to the controller.

Wiring for power

W A R N I N G

Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical
connections.

W A R N I N G

Potential Electrocution Hazard. If this equipment is used outdoors or in potentially wet locations, a
Ground Fault Interrupt device must be used for connecting the equipment to its mains power source.

English 11