Description

HOLD OUTPUTS

Sets the value the controller sends to an external system for a defined period of
time. After the time period, the instrument goes back to reporting real time
values.
ACTIVATION—LAUNCH or RELEASE
SET OUTMODE—HOLD OUTPUTS (default) or TRANSFER OUTPUTS
SET CHANNELS—ALL (default) or select from hardware list

TEST OUTPUT

• OUTPUT 1
• OUTPUT 2

Lets the user select a mA value that is sent by the controller for verification.
Min: 0 mA (default +04.00)
Max: 25.00 mA

STATUS

View status of all modules, sensors and relays.

TEST RELAY—A, B, C, D

ENERGIZE or DE-ENERGIZE the selected relay.

OVERFEED RESET

Resets the overfeed timer.

RESET DEFAULT
CONFIG

Resets the controller configuration settings to the default values (language, date
and time, relay function and data output function).

RESTART

Performs a controller restart.

SIMULATION (only
displays if sensors or
modules are connected)

After the SIM VALUE is entered, the controller outputs this value as if it was the
value sent from the sensor. The simulation stops after the user exits the screen.
SELECT SOURCE—

• <Module 1>
• <Module 2>

(footer displays current source selection)
SET PARAMETER—Source measurement type (footer displays current source
selection)
SET SIM VALUE—Use arrow keys to change value (footer displays current
source selection)

MODBUS STATS

Displays Error and Good count stats for selected port.

• Sensor port 1, 2, 3 or 4
• Network port
• Service port
• Clear stats

SYSTEM DATA

Displays the current system current, temperature and voltage data.

Sensor diagnostic and test menu

The sensor diagnostic and test menu shows current and historical information about the instrument.
Refer to

Table 6

.

Table 6 Sensor DIAG/TEST menu

Option

Description

MODULE INFORMATION

Shows information about the sensor module.

SENSOR INFORMATION

Shows the name and serial number that was entered
by the user.

CAL DAYS

Shows the number of days since the last calibration.

CAL HISTORY

Shows a list of all calibrations by date/time stamp. Use
the arrows keys to select a calibration and push enter
to view the details.

English 33