Table 7 Warning list for conductivity sensors (continued)

Warning

Description

Resolution

REPLACE SENSOR

The sensor has been in operation >
365 days

Calibrate the sensor with a
reference solution and reset the
sensor days. Refer to

Sensor

diagnostic and test menu

on page 33. If the calibration fails,
call technical support.

CAL IN PROGRESS

A calibration was started but not
completed

Return to calibration.

OUTPUTS ON HOLD

During calibration, the outputs were
set to hold for a selected time.

The outputs will become active after
the selected time period.

WRONG LINEAR TC

The user-defined linear temperature
compensation is out of range

The value must be between 0 and
4%/°C; 0 to 200 °C.

WRONG TC TABLE

The user-defined temperature
compensation table is out of range

The temperature is above or below
the temperature range defined by
the table.

WRNG USER CONC TABLE

The concentration measurement is
outside of the range of the user
table

Make sure that the user table is set
for the correct measurement range.

WRNG BLT-IN TEMP TABLE

The measured temperature is
outside of the range of the built-in
temperature compensation table

Make sure that the temperature
compensation is configured
correctly.

WRNG BLT-IN CONC TABLE

The concentration measurement is
outside of the range of the built-in
concentration table

Make sure that the concentration
measurement is configured for the
correct chemical and range.

Error list

Errors may occur for various reasons. An error icon consists of an exclamation point within a circle.
When an error occurs, the error icon and the measurement screen flash alternately in the main
display. All outputs are held when specified in the controller menu. To view errors, push the menu
key and select DIAGNOSTICS. Then select the device to view any problems associated with that
device.
A list of possible errors is shown in

Table 8

.

Table 8 Error list for conductivity sensors

Error

Description

Resolution

ADC FAILURE

The analog to digital conversion
failed

Make sure that the sensor module is
fully inserted into the controller
connector. Replace the sensor
module.

SENSOR MISSING

The sensor is missing or
disconnected

Examine the wiring and connections
for the sensor and for the module.
Make sure that the terminal block is
fully inserted into the module.

SENS OUT RANGE

The sensor signal is outside of the
accepted limits (2 S/cm)

Make sure that the display format is
set for the correct measurement
range.

Spare parts and accessories