Inductive Probe 8398

7

3.2 Compliance


i Probes are factory tested and are delivered with a certificate of compliance with

specifications.
Models 8398.2 and 8398.3 are particularly well-suited to applications where
sanitary requirements are extremely stringent (compliance with the European food
directive 90/128/CE modified by directive 93/9/CE). Indeed, these probes do not
possess any retention zones that could allow the development of bacteria
(interstices, slits, joints, etc.) and have a roughness inferior to 0.5 μm (optional
certificate, ref. 08398=T=1111).


i The measurement chain, made up of an 8398 probe and of our conductivity

transmitters, complies with European directives 89/336/CEE and 73/23/CEE
modified by directive 93/68/CEE.



4. PROBE

START-UP

4.1 Cable

connection


WARNING: Fit the DN50 clamping nut or the immersion rod onto the probe before

wiring!


Make the following connections

COND 9125

COND 8921
(channels 1 and 2)

1 : GREEN or yellow (Pt100)

TEMP +

14 and 19

2 : YELLOW or green (Pt100)

TEMP -

15 and 20

GND

30 and 32

3 : BLACK (secondary coil shielding)

4 : WHITE (secondary coil)

IN

31 and 33

GND

16 and 21

5 : BLACK (primary coil shielding)

6 : BROWN (primary coil)

OUT

17 and 22

7 : WHITE (external shielding)

EXTERNAL
SHIELDING

EXTERNAL SHIELDING


REMARK: The probe is fitted as standard with 5 metres of cable. We recommend

not cutting this cable in order to avoid any connection errors. If the
length is insufficient, please order the following accessories:

x Junction box, ref.: 08335=A=6000

x Cable (per metre, maximum 50 m), ref.: 150727,10000