Inductive Probe 8398

8

4.2 Use with a 8921 conductimeter






















The use of a type 8398 inductive probe connected to a type 8921 2-channel
conductimeter requires that this latter be fitted with a software version greater than or
equal to the following versions:

Software Minimum

version no.

Memory reference

Standard 3.76 08921=A=6200

Concentration 1.04

08921=A=6400

Special 4.07

08921=A=6305


To install your new software version, proceed as follows:
x Remove the transmitter's front panel (4 screws).

x On the rear of the front panel, there are two circuit boards. Remove the first of

these two circuit boards (4 screws).

x At the rear of this printed circuit board, carefully change the integrated circuit

(EPROM) represented on the figure opposite.

x Check that the memory has been inserted in the correct position, otherwise it

would be irreversibly damaged.

x Re-fit the front panel and initialise the instrument's default parameters using

command 900.

x Reprogram the instrument to your configuration and configure command C101

(102) with Argument 2, which corresponds to the 8398 probe's cell constant,
equal to 2.35.

EPROM