Hach-Lange POLYMETRON 8398 User Manual

Page 8

Advertising
background image

Inductive Probe 8398

5

2.2 Principle of inductive technology



1 : Receiver

2 : Oscillator

3 : Cables

4 : Thermoplastic insulator

5 : Receiver

6 : Transmitter

7 : Induced current




The 8398 probes are made up of two coils that are completely insulated from the
process:

x The primary (or transmitter) coil, supplied with alternating voltage, produces an

alternating electromagnetic field that generates an electrical current in the solution.


x The secondary (or receiver) coil detects the size of the weak current induced by

the movement of ions in the solution.



The absence of contact between the electrical part and the solution (magnetic
coupling) provides a large number of advantages compared to the traditional
technique using metallic electrodes:

x No polarisation and hence a broad measurement range.

x High chemical and mechanical resistance.

x Possibility of performing measurements in soiling products.

x Perfectly hygienic design.

Advertising