Inductive Probe 8398

9

4.3 Use with a 9125 conductimeter


Set the two switches (I/K) inside the instrument to position I.
Check that the measurement type is set to inductive.


5. PROBE

CALIBRATION


REMARK: For more detailed information, see the user instructions of out 9125 and

8921 transmitters. We recommend proceeding in the following order:

‰

Set the cell constant value on your transmitter to 2.35 cm

-1

.

‰

Calibrate your temperature sensor:

Submerge the probe in a solution for approximately 10 min., raise the temperature of
this solution using a thermometer (accurate to

r 0.1°C if possible). Set the transmitter

to process calibration mode, then adjust the temperature value.

‰

Calibrate conductivity (two solutions)

Electrical calibration:

Perform the "electrical" calibration of the probe with a decade box, using a wire no
longer than 50 cm and with a minimum cross-section of 0.38 mm².
The resistance values to use are:

f for zero adjustment (decade box disconnected):




then 200 Ohms (default instrument value) for slope adjustment: