Inductive Probe 8398

18


8. MAINTENANCE

WARNING: Read the "installation precaution" and "cable connection" chapters

carefully.

o

If there is a doubt concerning the operation of a probe, disconnect it from the

transmitter, leaving it in place on your process and ensure that you obtain the
following resistance values (see 4.1 Cable connection):



R (1-2)

| 110 : (at 25 °C) (Temperature sensor Pt100)

R (3-4)

| [1…2] : (Secondary coil)

R (5-6)

| [1…2] : (Primary coil)

R (3-5)

f : (Coil electrical insulation)

o

If the checks performed do not generate any erroneous resistance values, the

probe can be considered to be working properly.

o

If the measurement is still not satisfactory, remove the probe from the process and

calibrate it (see "probe calibration" chapter), set the transmitter to "no temperature
compensation", then check the following points:


Displayed conductivity (K = 2.35)

Simulated resistance

470 μS/cm

5 k

:

4.7 mS/cm

500

:

47 mS/cm

50

:

470 mS/cm

5

: