Hach-Lange pHD sc User Manual

Page 13

Advertising
background image

7

Section 3

Installation

DANGER
Only qualified personnel should conduct the tasks described in this section of the
manual.

3.1 Connecting/Wiring the Sensor to the sc Controller

3.1.1 Attaching a sc Sensor with a Quick-connect Fitting

The sensor cable is supplied with a keyed quick-connect fitting for easy attachment to the
controller (

Figure 5

). Retain the connector cap to seal the connector opening in case the

sensor must be removed. Optional extension cables may be purchased to extend the
sensor cable length. If the total cable length exceeds 100 m (300 ft), a termination box
must be installed.

Note: Use of a load termination box other than Cat. No. 5867000 may result in a hazard.

Figure 5

Attaching the Sensor using the Quick-connect Fitting

Figure 6

Quick-connect Fitting pin assignment

Number

Designation

Wire Color

1

+12 VDC

Brown

2

Circuit Common

Black

3

Data (+)

Blue

4

Data (–)

White

5

Shield

Shield (grey wire in existing quick-disconnect fitting)

6

Groove

1

2

3

4

5

6