Hach-Lange pHD sc User Manual
Page 13

7
Section 3
Installation
DANGER
Only qualified personnel should conduct the tasks described in this section of the
manual.
3.1 Connecting/Wiring the Sensor to the sc Controller
3.1.1 Attaching a sc Sensor with a Quick-connect Fitting
The sensor cable is supplied with a keyed quick-connect fitting for easy attachment to the
controller (
Figure 5
). Retain the connector cap to seal the connector opening in case the
sensor must be removed. Optional extension cables may be purchased to extend the
sensor cable length. If the total cable length exceeds 100 m (300 ft), a termination box
must be installed.
Note: Use of a load termination box other than Cat. No. 5867000 may result in a hazard.
Figure 5
Attaching the Sensor using the Quick-connect Fitting
Figure 6
Quick-connect Fitting pin assignment
Number
Designation
Wire Color
1
+12 VDC
Brown
2
Circuit Common
Black
3
Data (+)
Blue
4
Data (–)
White
5
Shield
Shield (grey wire in existing quick-disconnect fitting)
6
Groove
1
2
3
4
5