Specifications – Hach-Lange pHD sc User Manual

Page 8

Advertising
background image

2

Specifications

Temperature
Compensation

Automatic from –10 to 105 °C
(14.0 to 221 °F) with selection
for NTC 300 ohm thermistor, Pt
1000 ohm RTD, or Pt 100 ohm
RTD temperature element, or
manually fixed at a user-entered
temperature; additional
selectable temperature
correction factors (ammonia,
morpholine, or user-defined
pH/°C linear slope) available for
pure water automatic
compensation from 0.0 to 50 °C
(32 to 122 °F)

Automatic from –10 to 105 °C
(14.0 to 221 °F) with selection
for NTC 300 ohm thermistor, Pt
1000 ohm RTD, or Pt 100 ohm
RTD temperature element, or
manually fixed at a user-entered
temperature; additional
selectable temperature
correction factors (ammonia,
morpholine, or user-defined
pH/°C linear slope) available for
pure water automatic
compensation from 0.0 to 50 °C
(32 to 122 °F)

N/A

Measurement
Accuracy

±0.02 pH

±0.02 pH

±5 mV

Temperature Accuracy

±0.5 °C (0.9 °F)

±0.5 °C (0.9 °F)

±0.5 °C (0.9 °F)

Repeatability

±0.05 pH

±0.05 pH

±2mV

Sensitivity

±0.01 pH

±0.01 pH

±0.5 mV

Calibration Methods

Two point automatic, one point
automatic, two point manual,
one point manual.

Two point automatic, one point
automatic, two point manual,
one point manual.

one point manual

Maximum Probe
Immersion Depth/
Pressure

Submersible to 107 m (350
ft)/1050 kPa (150 psi)

Immersion only

Submersible to 107 m (350
ft)/1050 kPa (150 psi)

Sensor Interface

Modbus

Modbus

Modbus

Probe Cable Length

6 m (20 ft) + 7.7 m (25 ft)
interconnect cable extension for
analog sensor with digital
gateway

10 m (31 ft) for sensor with
integral digital electronics

6 m (20 ft) + 7.7 m (25 ft)
interconnect cable extension for
analog sensor with digital
gateway

10 m (31 ft) for sensor with
integral digital electronics

6 m (20 ft) + 7.7 m (25 ft)
interconnect cable extension
for analog sensor with digital
gateway

10 m (31 ft) for sensor with
integral digital electronics

Probe Weight

316 g (11 oz)

870 g (31 oz)

316 g (11 oz)

Probe Dimensions

See

Figure 2 on page 5

through

Figure 3 on page 5

.

See

Figure 4 on page 5

.

See

Figure 2 on page 5

through

Figure 3 on page 5

.

1

Most pH applications are in the 2.5 to 12.5 pH range. The pHD™ Differential pH sensor with the wide-range glass process

electrode performs exceptionally well in this range. Some industrial applications require accurate measurement and control
below 2 or above 12 pH. In these special cases, please contact the manufacturer for further details.

2

For best ORP measuring results in solutions containing zinc, cyanide, cadmium or nickel, the manufacturer recommends using

the pHD™ ORP sensor equipped with a gold electrode.

3

PEEK

®

is a registered trademark of ICI Americas, Inc.

4

Ryton

®

is a registered trademark of Phillips 66 Co.

5

Kynar

®

is a registered trademark of Pennwalt Corp.

6

Viton

®

is a registered trademark of E.I. DuPont de Nemours + Co.

Table 2 Digital Gateway Specifications

Weight

145 g (5 oz)

Dimensions

17.5 x 3.4 cm (7 x 1

3

/

8

in.)

Operating Temperature

–20 to 60 °C (–4 to 140°F)

Table 1 Differential pH and ORP Sensor Specifications (continued)

Specification Category

pH Sensors

1

Stainless Steel pH Sensor

ORP Sensors

2

Advertising