Hach-Lange pHD sc User Manual

Page 7

Advertising
background image

1

Section 1

Specifications

Specifications are subject to change without notice.

Table 1 Differential pH and ORP Sensor Specifications

Specification Category

pH Sensors

1

Stainless Steel pH Sensor

ORP Sensors

2

Wetted Materials

PEEK

®3

or Ryton

®4

(PVDF)

body, salt bridge of matching
material with Kynar

®5

junction,

glass process electrode, titanium
ground electrode, and Viton

®6

O-ring seals (pH sensor with
optional HF-resistant glass
process electrode has 316
stainless steel ground electrode,
and perfluoroelastomer wetted
O-rings; for other wetted O-ring
materials consult the
manufacturer)

Immersion mounting only,
316 SS Stainless Steel body
with Ryton

®

(PVDF) ends and

salt bridge.

PEEK

®

or Ryton

®

(PVDF)

body, salt bridge of matching
material with Kynar

®

junction,

glass and platinum (or glass
and gold) process electrode,
titanium ground electrode, and
Viton

®

O-ring seals

Operating Temperature
Range

–5 to 70 °C (23 to 158 °F) for
sensor with integral digital
electronics

–5 to 105 °C (23 to 221 °F) for
analog sensor with digital
gateway

0 to 50 °C (32 to 122 °F) for
sensor with integral digital
electronics

–5 to 70 °C (23 to 158 °F) for
sensor with integral digital
electronics

–5 to 105 °C (23 to 221 °F) for
analog sensor with digital
gateway

Pressure/Temperature
Limits
(without mounting
hardware)

6.9 bar at 105 °C (100 psi at
221 °F) for analog with gateway
6.9 bar at 70 °C (100 psi at
158 °F)

N/A (immersion only)

6.9 bar at 70 °C (100 psi at
158 °F)

6.9 bar at 105 °C (100 psi at
221 °F) for analog with
gateway

Maximum Flow Rate

3 m (10 ft) per second

3 m (10 ft) per second

3 m (10 ft) per second

Built-in Temperature
Element

NTC 300 ohm thermistor for
automatic temperature
compensation and analyzer
temperature readout

NTC 300 ohm thermistor for
automatic temperature
compensation and analyzer
temperature readout

NTC 300 ohm thermistor for
analyzer temperature readout
only — not for automatic
temperature compensation

Stability

0.03 pH per 24 hours,
non-cumulative

0.03 pH per 24 hours,
non-cumulative

2 mV per 24 hours,
non-cumulative

Maximum
Transmission Distance

1000 m (3280 ft) with
termination box

1000 m (3280 ft) with
termination box

1000 m (3280 ft) with
termination box

Sensor Cable (integral)

Digital: PUR (polyurethane)
4-conductor with one shield,
rated to 105 °C (221 °F), 10 m
(33 ft) standard length

Analog: Five-conductor (plus two
isolated shields) cable with
XLPE (cross-linked
polyethylene) jacket; rated to
150 °C (302 °F); 6 m (20 ft)
standard length

Digital: PUR (polyurethane)
4-conductor with one shield,
rated to 105 °C (221 °F), 10 m
(33 ft) standard length

Digital: PUR (polyurethane)
4-conductor with one shield,
rated to 105 °C (221 °F), 10 m
(33 ft) standard length

Analog: Five-conductor (plus
two isolated shields) cable
with XLPE (cross-linked
polyethylene) jacket; rated to
150 °C (302 °F);
6 m (20 ft) standard length

Components

Corrosion-resistant materials,
fully-immersible probe with 10 m
(30 ft) cable

Corrosion-resistant materials,
fully-immersible probe with 10 m
(30 ft) cable

Corrosion-resistant materials,
fully-immersible probe with 10
m (30 ft) cable

Measuring Range

–2.0 to 14.0 pH or –2.00 to 14.00
pH

–2.0 to 14.0 pH or –2.00 to
14.00 pH

–1500 to +1500 mV

Probe Storage
Temperature

4 to 70 °C (40 to 158 °F); 0 to
95% relative humidity,
non-condensing

4 to 70 °C (40 to 158 °F); 0 to
95% relative humidity,
non-condensing

4 to 70 °C (40 to 158 °F); 0 to
95% relative humidity,
non-condensing

Advertising