1 error codes, 2 warnings – Hach-Lange pHD sc User Manual

Page 29

Advertising
background image

23

Section 6

Troubleshooting

6.1 Error Codes

Errors are defined in

Table 4

.

6.2 Warnings

Errors are defined in

Table 5

.

Table 4 Error Codes

Displayed Error

Definition

Resolution

ADC FAILURE

System measurement fails

Contact Technical Consulting Services.

Table 5 Warning Codes

Displayed Warning

Definition

Resolution

PROBE OUT RANGE

Measured pH/ORP exceeds the expected
value range.

Contact Technical Consulting Services.

TEMP OUT RANGE

Measured temperature exceeds the
expected value range.

Contact Technical Consulting Services.

FLASH FAILURE

System flash memory write has failed.

Contact Technical Consulting Services.

ACTIVE. ELEC

Standard electrode is not performing within
the required specifications.

Contact Technical Consulting Services.

REF. ELECTRODE

Reference electrode is not performing
within the required specifications.

Contact Technical Consulting Services.

CAL REQUIRED

60 days has elapsed since the last
calibration

Perform a calibration.

REPLACE SENSOR

One year has elapsed since the sensor has
been installed.

Replace sensor and restore counter in the
SENSOR SETUP>DIAG/TEST>RESET SENSOR
menu