14

Operation

TEMP ADJUST

Adjust the displayed temperature by up to ± 15 °C.

DEFAULT SETUP

Restores the system to the original factory calibration.

CONFIGURE

EDIT NAME

Enter up to a 10-digit name in any combination of symbols and alpha or numeric characters.

SELECT MEASURE

Select the appropriate measurement units to display.

DISPLAY FORMAT

Select the measurement resolution (xx.xx pH or xx.x pH).

TEMP UNITS

Choose from the displayed options (°C or °F).

LOG SETUP

Choose SENSOR INTERVAL to set the sensor log interval or select TEMP INTERVAL to set the temperature log interval.

REJECT FREQ

Choose 50 or 60 Hz depending on the power line frequency for optimal noise rejection. Default is 60 Hz.

FILTER

Select 0–60 second signal averaging time.

TEMP ELEMENT

Select type of temperature element from the displayed choices.

SELECT BUFFER

Select the buffer type (standard 4, 7, 10 or DIN 19267) from the displayed choices.

PURE H20 COMP

Allows the user to specify that ammonia, morpholine, or other user-defined electrolyte is being used in the application,
allowing a temperature-dependent linear slope factor to be applied to the measured pH.

CAL DAYS

Number of days since the last calibration. Default notification at 60 days.

SENSOR DAYS

Number of days the sensor has been in operation. Default notification at 365 days.

DEFAULT SETUP

Resets all user-editable options to their factory-defaults.

DIAG/TEST

PROBE INFO

Display the sensor type, entered name of the sensor (Default: sensor serial number.), the sensor serial number, the
software version number, and the sensor driver version number.

CAL DATA

Displays the pH slope and the date of the last calibration.

4.5 pH Sensor Setup Menu (continued)