16

Operation

Resets all user-editable options to their factory-defaults.

CONFIGURE

EDIT NAME

Enter a 10-digit name in any combination of symbols and alpha or numeric characters.

SELECT MEASURE

Choose from Conductivity, Concentration, TDS, or Salinity. If Concentration is selected, the option to configure the
concentration settings is offered. Press the down arrow to Config Conc. Two concentration types are offered: Built-in and
User Defined. The Built-in menu offers an array of chemicals and concentrations to choose from.

COND UNITS

Choose from the displayed units (dependent on the parameter selected in the Select Measure menu)

TEMP UNITS

Select Celsius or Fahrenheit.

T–COMPENSATION

The factory default for temperature compensation is linear with a 2.00% per °C slope and a 25 °C reference temperature.
The default settings are appropriate for most aqueous solutions. The menu options are different depending on the
measurement that is selected in the Configure>Select Measure menu. The options are as follows:

NONE: No temperature compensation.

LINEAR: Recommended for most applications. Press

ENTER

to change the slope or reference temperature.

NATURAL WATER: Contact Technical Consulting Services for application specific information and assistance.

TEMP TABLE: Use to configure a temperature compensation table by entering up to 10 x-axis values and 10 y-axis values.
See

section 4.8.1 on page 21

to enter values. Contact Technical Consulting Services for additional information and

assistance.

LOG SETUP

Choose from Sensor Interval or Temp Interval. If the interval is enabled, choose from the displayed options to specify the
frequency to log the sensor or temperature reading. Default is Disabled.

AC FREQUENCY

Select 50 Hertz or 60 Hertz to specify the frequency to be excluded.

FILTER

Average the measurement over time by entering a number between 0–60. Default is 0 seconds.

TEMP ELEMENT

Select the temperature element type (100PT, 1000PT (default), or manual) then choose Select Factor to enter the specific
“T” Factor supplied with the sensor.

CAL DAYS

Number of days since the last calibration. Default notification at 60 days.

SENSOR DAYS

Number of days the sensor has been in operation. Default notification at 365 days.

DEFAULT SETUP

Reset the configure settings to the factory defaults.

4.5 Sensor Setup Menu (continued)