Operation

4.6 Pressure and Elevation

Note: If the barometric pressure from

Table 5

is entered in the meter, the altitude entered in

combination with this value must be 0 feet.

Table 5

can be used to estimate the true barometric pressure at certain elevations. The

correspondence is based on the assumption that at sea level the barometric pressure is
760 mm Hg. After determining the barometric pressure from the table or obtaining it from a
local weather service, enter this value into the instrument.

DIAG/TEST

PROBE INFO

Display the entered name of the sensor, the sensor serial number, the software version number, and the sensor driver
version number.

CAL DATA

Displays the current offset correction and the date of the last calibration.

SIGNALS

SENSOR SIGNAL allows the user to set the sensor range and display the sensor ADC counts and TEMP ADC COUNTS
shows raw data for temperature ADC counts. Comparable to A/D counts.

COUNTERS

SENSOR DAYS displays the cumulative days the sensor has been in use and RESET SENSOR allows the sensor counter
to be reset to zero.

4.5 Sensor Setup Menu (continued)

Table 5 Elevation Barometric Pressure

Elevation in feet

Barometric pressure in mm Hg

Elevation in feet

Barometric pressure in mm Hg

0

760

6000

613

500

746

6500

601

1000

733

7000

590

1500

720

7500

579

2000

708

8000

568

2500

695

8500

559

3000

683

9000

548

3500

671

9500

538

4000

659

10000

527

4500

647

10500

517

5000

635

11000

506

5500

624