Operation 4.7.4 cond cal, 5 concurrent calibration of two sensors – Hach-Lange 3705 sc User Manual

Page 22

Advertising
background image

20

Operation

4.7.4 Cond Cal

1.

From the Main Menu, select SENSOR SETUP and confirm.

2.

Highlight the appropriate sensor if more than one sensor is attached and confirm.

3.

Select CALIBRATE and confirm.

4.

Select COND CAL. Select the available Output Mode (Active, Hold, or Transfer) from
the list box and confirm.

5.

Select SET REF TEMP and confirm.

6.

Select SET SLOPE and confirm.

7.

Move the probe to the solution and confirm to continue.

8.

Confirm when stable. A CAL Complete will be displayed and confirm.

9.

Return the sensor to process.

4.7.5 Concurrent Calibration of Two Sensors

1.

Begin a calibration on the first sensor and continue until “Wait to Stabilize”
is displayed.

2.

Select LEAVE and confirm. The display will return to the Main Measurement screen
and the reading for both sensors will be flashing.

3.

Begin the calibration for the second sensor and continue until “Wait to Stabilize”
is displayed.

4.

Select LEAVE. The display will return to the Main Measurement screen and the
reading for both sensors will be flashing. The calibration for both sensors are now
running in the background.

5.

To return to the calibration of either sensor select the Main Menu. Select Sensor Setup
and confirm. Select the appropriate sensor and confirm.

6.

The calibration in progress will be displayed. Continue with the calibration.