5.5 The commands under SENSOR SETUP

SENSOR SETUP

SELECT SENSOR (for several sensors)

CALIBRATE

ZERO CAL

Eliminates sensor offset
See

5.6.1 "Calibration in air (ZERO CAL)"

.

ELECTRIC. SPAN

Calibration with a defined resistance.
See

5.6.2 "Calibration in air (ELECTRIC. SPAN)"

.

PROCESS SPAN

Calibration with a reference solution.
See

5.6.3 "Calibration in the process (PROCESS SPAN)"

.

PROCESS TEMP

Calibration of the temperature.
See

5.7 "Adjusting the Temperature"

.

CAL CONFIG

OUTPUT_MODE

Choose between:
ACTIVE, output signal follows the input signal;
HOLD, last measured value and output signal are held;
TRANSFER, fixed value is output to the peripherals; and
CHOICE

CAL REMINDER

You can set when the next calibration is to be performed. The
controller then automatically indicates when the next calibration is due.

CORR FACTORS

Choose between
TEMP OFFSET,
GAIN CORR and
GAIN VALUE,

SET CAL DEFLT

Return to the factory settings after prompt for confirmation.

CONFIGURE

EDIT NAME

Enter a 10-character name.

PARAMETERS

Choose a parameter.

DEGREES C-F

Choose between degrees Celsius or Fahrenheit.

T-COMPENSATION

Choose between LINEAR and NONE.

T-SENSOR

AUTOMATIC
MANUAL

Enter a value.

FILTER

Enter a value.

LOG. DELAY

LOG INTERVAL

Choose between the values available or DISABLED.

TEMP INTERVAL

Choose between the values available or DISABLED.

MAINS FREQ.

Enter the mains frequency.

SET DEFAULTS

Returns to the factory settings after a prompt for confirmation.

DIAG/TEST

PROBE INFO

Provides information on driver, software and serial number

CAL DATA

Provides information on GAIN, Offset (T) and GAIN CORR.

SIGNALS

Provides information on raw data measured.

COUNTERS

Provides information on the operating time since the last calibration

TEST/MAINT

Disable OUTPUT during test or maintenance

SERVICE

Reset Service Counter