5.6 Sensor calibration (conductivity)

The sensor has been permanently calibrated and operates with great enough
precision and stability that calibration is rarely necessary.

Calibrate the sensor

as required (measured value outside the permitted tolerance,
see Section

7.3 "Important service data"

) or

in accordance or agreement with the authorities.

5.6.1 Calibration in air (ZERO CAL)

1. On the controller, open the menu SENSOR SETUP => CALIBRATE => ZERO

CAL and accept.

2. Remove the sensor from the tank, clean and dry.

3. Accept.

4. Wait until the controller displays CAL COMPLETE.

5. Confirm that you have replaced the sensor in the sample flow and accept.

5.6.2 Calibration in air (ELECTRIC. SPAN)

1. Remove the probe from the flow of sample, clean and dry.

2. Connect a defined resistance (in range 5 Ohms - 5 kOhms) and accept to

continue.

3. Enter the value for the resistance and accept.

4. Wait until the value has stabilised.

5. Accept the value.

6. Replace the probe in the flow of sample and accept.

5.6.3 Calibration in the process (PROCESS SPAN)

1. Hold the cleaned probe in the solution and accept to continue.

2. Accept when the measured value has stabilised.

3. Enter the value and accept.

4. Replace the probe in the flow of sample and accept.